All electrical equipment and installations need to be safe for users and the same systems. Electricity conductors need to be insulator from each other, so as not to create danger or unnecessary consumption. A poorly insulated circuit can create leakage currents that can be dangerous and trigger your GFCI, RCCB or ELCB. Each country regulates these levels, where isolation is acceptable. Check the continuity of the cables, circuit connections, contact closure, the so-called "circuit breakers", fuses, connection resistance of the connections etc. are very important measures.
Large Dual-LCD with graphic bar and backlight
MIN / MAX recording and Data-Hold
Auto discharge of capacitive voltages for user protection
Active safety circuit prevents the test if the voltage found is greater than 30 V
Low battery indicator
Auto power off
Safety: TÜV / GS, EN-61010-1; CAT III 1000 V
Accessories: case, batteries, test probes and manual
Insulation Resistance: • Test voltage: 125/250/500/1000 V; + 10% / -0%
Test currents: •• 1 mA with 125 kOhm at 125V
1 mA with 250 kOhm at 250V
1 mA with 500 kOhm at 500V
1 mA with 1 MOhm at 1000V
AC voltage: 750 V - 1 V; ± 1.2% + 10 dgt.
DC voltage: 1000 V - 1 V; ± 0.8% + 3 dgt.
Resistance: 40/400 Ohms - 10 mOhms; ± 1.2% + 3 dgt.
Power supply: 6 x 1.5 V AA (UM-3) batteries
Dimensions: 92 x 200 x 50 mm
Weight: 700 g